INITIAL PREPARATION
Each meter is supplied complete with a 9 V battery. Remove the back
cover, unwrap the battery and install it while paying attention to its
polarity its (see "Battery Replacement" section).
Connect the probe to the meter securely by
aligning the pins with the socket, pushing
the plug in and tightening the threaded ring.
Make sure that the sleeve is properly inserted on the probe, with the
holes towards the top of the probe (i.e. the end nearest to the cable).
Make sure that the meter has been calibrated before taking any
measurements (see "Calibration" section for details).
TAKING MEASUREMENTS
To take measurement, place the probe into
the solution to be tested while making sure
that the holes are completely submerged.
Tap and stir the probe to remove any air
bubbles trapped inside the PVC sleeve.
Turn the instrument on by pressing the
ON/OFF key, and then select the desired
range by pressing the corresponding key.
If the LCD displays only a "1" on the far left
hand side, the meter is out of range and
the solution measured exceeds the selected
range. In this case, the next (higher) range
should be selected.
OPERATIONAL GUIDE
ON/OFF
3
4
2
1
FUNCTIONAL DESCRIPTION OF
HI 76302W CONDUCTIVITY PROBE
1) Watertight screened cable (1 m / 3.3')
2) 4 stainless steel rings
3) Air-release holes
4) PVC protective sleeve