Hanna-instruments HI 96727 User Manual

Browse online or download User Manual for Sensors Hanna-instruments HI 96727. Hanna Instruments HI 96727 User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 2
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
1. GLP/ key: press to enter
GLP mode
. In
calibration mode
press to
edit the date and time.
2. CAL CHECK key: press to perform the validation of the meter, or
press and hold for three seconds to enter
calibration mode
.
3. ZERO/CFM key: press to zero the meter prior to measurement, to
confirm edited values or to confirm factory calibration restore.
4. READ/ key: In
measurement mode
, press to make a measurement, or
press and hold for three seconds to start a pre-programmed countdown
prior to measurement. In
GLP mode
press to view the next screen.
5. ON/OFF key: to turn the meter on and off.
6. Liquid Crystal Display (LCD)
7. Cuvette alignment indicator
8. Cuvette holder
DISPLAY ELEMENTS DESCRIPTION:
INSTRUCTION MANUAL
HI 96727HI 96727
HI 96727HI 96727
HI 96727
www.hannainst.com
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is
not damaged. If any damage occured during shipment, please notify
your Dealer.
Each HI 96727 Ion Selective Meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 9V Battery
• Instruction Manual
Note: save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.
For more details about spare parts and accessories
see “Accessories”.
Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you
with the necessary information for the correct use of the instrument. Please
read it carefully before using the meter. If you need additional technical
information, do not hesitate to e-mail us at [email protected].
Functional description:
Light High: There is too much light to perform
a measurement. Please check the preparation of
the zero cuvette.
Light Low: There is not enough light to perform
a measurement. Please check the preparation of
the zero cuvette.
No Light: The instrument cannot adjust the
light level. Please check that the sample does
not contain any debris.
Inverted cuvettes: The sample and the zero
cuvette are inverted.
Zero: A zero reading was not taken. Follow the
instructions of the measurement procedure for
zeroing the meter.
Under range: A blinking “0” indicates that
the sample absorbs less light than the zero
reference. Check the procedure and make sure
you use the same cuvette for reference (zero)
and measurement.
Over Range: A flashing value of the maximum
concentration indicates an over range condition.
The concentration of the sample is beyond the
programmed range: dilute the sample and
re-run the test.
Standard Low: The standard reading is less
than expected.
Standard High: The standard reading is higher
than expected.
Errors and warnings:
Measurement procedure:
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.
3• Fill one cuvette up to the 10 mL mark
with deionized water and replace the cap.
This is the blank.
4• Fill another cuvette up to the 10 mL mark
with unfiltered sample and replace the
cap. This is the apparent color.
5• Filter 50 mL of sample through a filter
with a 0.45 µm membrane in a beaker.
6• Fill the third cuvette up to the 10 mL
mark with the filtered sample and replace
the cap. This is the true color.
7• Place the blank into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
8• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
9• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for measurement. Remove the cuvette.
10•Insert the apparent color cuvette into the
instrument and ensure that the notch on
the cap is positioned securely into the
groove.
11•Press READ/ and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
12•The meter directly displays the value of
apparent color.
13•Remove the cuvette, insert the true color
cuvette into the instrument and ensure
that the notch on the cap is positioned
Measurement
1. The measuring scheme (lamp, cuvette, detector), appears during
different phases of zero or reading measurement
2. Error messages and warnings
3. The battery icon indicates the charge state of the battery
4. The hourglass appears when an internal check is in progress
5. Status messages
6. The chronometer appears when the reaction timer is running
7. The month, day and date icons appear when a date is displayed
8. Four digit main display
9. Measuring units
10. Four digit secondary display
ON ZERO READING:
ON SAMPLE READING:
DURING CALIBRATION PROCEDURE:
OTHER ERRORS AND WARNINGS:
Cooling lamp: The instrument waits for the
lamp to cool down.
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery
is dead and must be replaced. Once this
indication is displayed, normal operation of the
instrument will be interrupted. Change the
battery and restart the meter.
Cap error: Appears when external light enters
in the analysis cell. Assure that the cuvette cap
is present.
Color of Water ISM
2
8
9
10
5
11
12
3-4
10 mL
6
10 mL
7
13
Technical specifications:
Range 0 to 500 PCU (Platinum Cobalt Units)
Resolution 10 PCU
Accuracy ±10 PCU ±5% of reading @ 25°C
Typical EMC Dev. ±10 PCU
Light Source Lamp and filter @ 420 nm
Light Detector Silicon Photocell
Method Adaptation of the Standard Methods for
the Examination of Water and Wastewater
18
Th
edition, Colorimetric Platinum Cobalt
method.
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode;
with last reading reminder.
Dimensions 192 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
Weight 360 g (12.7 oz.).
Page view 0
1 2

Summary of Contents

Page 1 - Color of Water ISM

1. GLP/ key: press to enter GLP mode. In calibration mode press toedit the date and time.2. CAL CHECK key: press to perform the validation of the met

Page 2 - CALIBRATION

14•Press GLP/ to edit the desired month(01-12). If the key is kept pressed, themonth number is automatically increased.15•When the correct month has