Hanna-instruments HI 96725 User Manual

Browse online or download User Manual for Sensors Hanna-instruments HI 96725. Hanna Instruments HI 96725 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - HI 96725C

1www.hannainst.comInstruction ManualHI 96725CChlorine, Cyanuric Acid & pH ISM

Page 2 - TABLE OF CONTENTS

18 19The reading should be within specifications as reported in the CAL CHECKTM StandardCertificate. If the value is found out of the specifications,

Page 3 - PRELIMINARY EXAMINATION

20 21• Remove the cuvette.• Place the specific CAL CHECKTM StandardCuvette B into the cuvette holder, for:Free Chlorine: B, HI 96701-11Total Chlorine:

Page 4 - PRECISION AND ACCURACY

22 23To display the calibration date:• Press and hold for three secondsRANGE/GLP/ to enter GLP mode. Thecalibration month and day will appear onthe m

Page 5 - FUNCTIONAL DESCRIPTION

24 25ACCESORIESREAGENT SETHI 93701-01 Reagents for 100 Free Chlorine testsHI 93701-03 Reagents for 300 Free Chlorine testsHI 93710-01 Reagents for 100

Page 6 - ERRORS AND WARNINGS

26 27Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environ

Page 7

28Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe

Page 8 - RANGE SELECTION

2 3TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 4GENERAL

Page 9 - MEASUREMENT PROCEDURE

4 5GENERAL DESCRIPTIONABBREVIATIONSThe HI 96725 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a

Page 10 - VALIDATION PROCEDURE

6 7PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cro

Page 11 - CALIBRATION PROCEDURE

8 91) RANGE/GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cu

Page 12

10 11ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are also displayedwhen the obtained values are

Page 13 - BATTERY REPLACEMENT

12 13The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts ma

Page 14 - WARRANTY

14 15• Press RANGE/GLP/ to change the range.The range can be changed any time when theinstrument is in measurement mode.The selected range is memoriz

Page 15 - (ROMANIA)

16 17• After a few seconds, the display willshow “-0.0-”. The meter is now zeroedand ready for measurement.• Remove the cuvette.• Add the specific tes