EC/TDS CALIBRATION
EC/TDS calibration is a single point procedure. Selectable calibration
points are: 0.0, 84.0 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88
mS/cm, 80.0 mS/cm,111.8 mS/cm.
To enter EC calibration, select the EC range and
press the CAL key.
Note: TDS reading is automatically derived from the EC reading and
no specific calibration for TDS is needed. Pressing CAL while TDS
range is selected has no effect.
Rinse the probe with some of the calibration solution or
deionized water. Immerse the probe into the solution.
The sleeve holes must be completely submerged. Tap
the probe repeatedly to remove any air bubbles that
may be trapped inside the sleeve.
For zero calibration, just leave the dry probe in air.
The indications "BUF" and "CAL" are dis-
played. The upper LCD shows the uncalibrated
EC reading. The lower LCD shows the buffer
value. The stability indicator " " blinks.
Select the desired value with the
and keys, if necessary.
When the " " symbol stops blinking, the
reading is stable. The "CON" indication starts
blinking on the LCD asking for confirmation.
Press ALT + CFM to confirm the calibration.
If everything is satisfactory, the meter displays the "Stor Good"
message and returns to measurement mode.
Notes
I If the uncalibrated reading is too far from the expected value,
calibration is not recognized. The "CON" indication does not
appear; the " " and "BUF" symbols blink to warn the user.
II For best results choose a calibration solution with an EC value
as close as possible to the sample to be measured.
TEMPERATURE COMPENSATION
Three options of compensating temperature are available:
Automatic (Atc): The probe has a built-in temperature sensor; the
value of the temperature is used to automatically compensate the EC/
TDS reading using 25°C as reference temperature. This is the default
option.
Manual (Mtc): The temperature value, shown on the lower LCD, can
be manually set by the user with the up and down arrow keys. The
compensation is referenced at 25°C. The "°C" symbol blinks when this
option is active.
No Compensation (notc): The temperature reading shown on the
lower LCD is not taken into account. The reading displayed on the
upper LCD is the actual EC or TDS value. The "°C" and "%TC" symbols
blink when this option is active.
To select the desired option press the ATC key until
the option is briefly displayed on the LCD.
Note: The default compensation mode is ATC.
If temperature compensation is active, measurements are compen-
sated using a default temperature coefficient of 1.90 %/°C.
It is possible to select a different temperature coefficient (TC) in the
0.00 to 6.00 %/°C range by entering the setup mode and selecting
the "tc" item (see SETUP for details).
The current temperature coefficient can be
quickly viewed pressing ALT+TC. The value
is briefly displayed on the lower LCD.
The EC and TDS scales are autoranging. The meter automatically sets
the scale with the highest possible resolution.
By pressing ALT+FIXED, the autoranging
feature is disabled and the current range
is frozen on the LCD. "F1" symbol blinks
on the LCD.
To restore the autoranging option press ALT+FIXED again.
Note: Autoranging is automatically restored if the RANGE key is pressed,
if the setup or calibration modes are entered and if the meter is
turned off and back on again.