Hanna-instruments HI 95707 User Manual

Browse online or download User Manual for Measuring instruments Hanna-instruments HI 95707. Hanna Instruments HI 95707 User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 10
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
120
MAN95707CR3 06/06
www.hannainst.com www.hannainst.com
This Instrument is in
Compliance with the CE Directives
Instruction Manual
HI 95707C
Nitrite Low Range ISM
Page view 0
1 2 3 4 5 6 7 8 9 10

Summary of Contents

Page 1 - Nitrite Low Range ISM

120MAN95707CR3 06/06www.hannainst.com www.hannainst.comThis Instrument is inCompliance with the CE DirectivesInstruction ManualHI 95707CNitrite Low

Page 2 - CE DECLARATION OF CONFORMITY

10 11Under range. A blinking "0.000" indicates that the sample absorbsless light than the zero reference. Check the procedure and makesure y

Page 3 - PRELIMINARY EXAMINATION

2 19PRELIMINARY EXAMINATION ... 3GENERAL DESCRIPTION...

Page 4 - GENERAL DESCRIPTION

318REAGENT SETSHI 93707-01 Reagents for 100 Nitrite Low Range testsHI 93707-03 Reagents for 300 Nitrite Low Range testsOTHER ACCESSORIESHI 95707-11 CA

Page 5 - CALIBRATION PROCEDURE

4 17The HI 95707 is an auto diagnostic portable microprocessor meter that benefits from Hanna’s years ofexperience as a manufacturer of analytical ins

Page 6 - PRINCIPLE OF OPERATION

516Range 0.000 to 0.600 mg/L (as NO2–-N)Resolution 0.001 mg/LPrecision ±0.001 mg/L @ 0.100 mg/LTypical EMC Deviation ±0.001 mg/LLight Source Tungsten

Page 7 - FUNCTIONAL DESCRIPTION

6 15• Turn the meter on by pressing ON/OFF.When the LCD displays “707”, it is ready.• Place the CAL CHECK™ Standard Cuvet A intothe holder and ensure

Page 8 - MEASUREMENT PROCEDURE

714A microprocessor controlled special tungsten lamp emits radiation which is first optically conditionedand beamed to the sample contained in the cuv

Page 9 - GUIDE TO DISPLAY CODES

8 13KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO: to zero the meter prior to measurement.• READ/TIMER: this is a bi-functional key,

Page 10

912This prompt appears for 1 second each time the instrument is turnedon.This indicates that the meter is in a ready state and zeroing can beperformed

Related models: HI 95707C

Comments to this Manuals