Hanna-instruments HI 93703 User Manual Page 7

  • Download
  • Add to my manuals
  • Print
  • Page
    / 13
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 6
1312
CALIBRATION
A monthly calibration is recommended. To
check the date of last calibration, simply hold
the DATE/ key down for few seconds.
A more frequent check of the instrument is
suggested by using the supplied standard solu-
tion.
CALIBRATION PROCEDURE
Turn the meter on and
wait for the display to
show "----".
Press the CAL key
once, the "CAL" mes-
sage will blink on the
display for about 6
seconds, then the
calibration mode
stops.
While the "CAL" mes-
sage is still blinking,
press CAL again. The
instrument is now in
the calibration mode
and a "CL" will ap-
pear on the lower part
of the display. The
date of calibration can
be edited now by sim-
ply pressing the
DATE/ key. To
scroll to the correct
number press the
READ/key. The de-
fault blinking param-
eter is the month, on
the left hand of the
display (MM.DD).
SOURCES OF INTERFERENCE
Presence of floating debris and coarse sedi-
ments which settle out rapidly will give false
readings.
The infrared light source used for HI 93703
turbidity meter, according to ISO 7027 Inter-
national Standard, can effectively minimize
errors due to colored dissolved substances.
This effect, named "true color", is a common
interference for most commercially available
instruments operating in the range of visible
light.
Air bubbles and the effect of vibrations that
disturb the surface of the sample will give
false results.
Dirty glassware could also affect readings
along with scratched or edged vials.
Page view 6