TAKING MEASUREMENTS
For a rapid response, expose the end of the
humidity detector to an air current moving at
more than 0.5 m/second.
In absence of air movement, the response
can be speeded up by shaking the probe.
If condensed drops form upon the surface of
the humidity sensors, turn the instrument off
and wait until this undesired water has com-
pletely evaporated. To speed up the evapora-
tion process, expose the humidity sensors to
an air current.
If the meter is used in dusty surroundings, the
filter must be kept on at all time.
Calibration procedure
Bring the calibration kit to a temperature of
approximately 20 ºC.
Remove the plug from the chamber contain-
ing the LiCl solution and insert the probe.
Remove the adhesive sticker which covers
the calibration trimmer access holes.
Wait for the reading to stabilize (this will take
a few hours) and then turn the "Low humidity
trimmer" (the one nearest to the probe con-
nection cable) until the value of 11.1% RH will
be displayed on the LCD.
Remove the probe and tightly seal the cham-
ber containing the LiCl solution.
Remove the plug from the second chamber
and insert the probe.
Wait for the reading to stabilize and then note
the displayed value.
By using the Calibration Table (pages 8-9),
find the displayed readout value on column
"A" and the corresponding RH value of col-
umn "B", to which the meter has to be brought
by acting on the "High humidity trimmer".
Adjust the "High humidity trimmer" (the one
opposed to the connection cable) until the
required value is displayed on the LCD.
Finally adjust the "Low humidity trimmer" until
the value of 75.4% RH will be displayed on the
LCD.
The HI 8064 thermohygrometer is factory cali-
brated and therefore usually it requires no fur-
ther intervention by the user.
If recalibration is desired, proceed as follows:
Remove the adhesive sticker which covers the
calibration trimmer access holes.
Insert the probe into a recipient in which air
heated at a temperature of 50 ºC has been
forced to circulate.
Insert a thermometer with an accuracy of 0.1
ºC into the same recipient.
Read the temperature value on the thermom-
eter.
Adjust the temperature calibration trimmer
(the central trimmer) until the measured tem-
perature value is displayed on the LCD.
TEMPERATURE CALIBRATION